Frequently-Asked Questions

Frequently-Asked Questions
How do I choose the right gas analyzer for a particular application? What factors need to be considered?
  • Confirm need gas analyzer continuous 24Hour*7 Days monitoring; measure one sampling point or inspect multiple sampling points, we can recommend you to use online or portable gas analyzer.
  • The concentration range of the measured gas and other background gases.
  • Other special function requirements, such as: output 4-20mA, RS485,RS232, or whether to use a printer or an industrial computer for data conversion, etc.
  • We will provide an on-site working condition table to understand measured gas, moisture, dust, tar, temperature, pressure, etc.
What are the types of common gas analyzers? What's the difference between them?

According to the technical principle, the commonly used is NDIR gas analyzer, UV-DOAS gas analyzer, TDLAS gas analyzer ,GC-FID gas analyzer,FTIR gas analyzer. The same gas can be measured by many different technologies. We can choose the appropriate gas analyzer for customers according to each customer applications.

such as H2S gas

  • Due to the high analytical sensitivity of the lead acetate paper strip colorimetric method, the hydrogen sulfide content can be detected at 10×10. This type of instrument is mainly used for the monitoring of trace H2S, and the measurement range is narrow. When detecting samples with H2S content greater than 50umol/mol, the instrument needs to add a dilution system, resulting in dilution measurement errors. In addition, the lead acetate paper tape is a consumable product, which needs to be replaced frequently, and the operating cost is relatively high.
  • The ultraviolet absorption method adopts the characteristic spectral absorption principle of SO2, H2S and other gases in the ultraviolet spectral region, with high detection sensitivity and accuracy and wide measurement range. There is almost no moisture interference in the measurement of SO2, H2S and other gases by ultraviolet absorption method, which is suitable for online detection of sulfide with high humidity in the measured gas.
  • The laser optical method uses the tunable diode laser absorption spectroscopy (TDLAS) measurement technology, which is commonly used for H2S monitoring in natural gas. The working wavelength of H2S measured by laser spectroscopy is 1577nm in the near-infrared region, and the lower detection limit can reach 5umol/mol per optical path, but it is easily interfered by gases such as CH4 and CO2, and the price of mid-infrared gas analyzers is high.
  • On-line gas chromatography Process gas chromatography can be equipped with various chromatographic detectors according to monitoring needs, so as to be suitable for on-line monitoring of different gases. It has the characteristics of many measurement components, wide measurement range, low detection limit and good repeatability. Thermal conductivity detectors (TCD) are mostly used for constant hydrogen sulfide gas detection, and flame photometric detectors (FPD) are mostly used for trace hydrogen sulfide gas detection.
  • The electrochemical sensor method is widely used in the detection and alarm of toxic and harmful gases. Electrochemical sensors are dedicated to the detection of SO2, H2S and other sulfur-containing gases. Although the measurement accuracy is not high, the price is low and easy to use. Most of them are used for portable SO2, H2S and other gas detectors, as well as for poisonous and harmful gases. Fixed detection and alarm system.
Delivery method and cycle of gas analysis equipment?

Delivery method and cycle of gas analysis equipment?

What gas components can a gas analyzer measured?

At present, the gases we can measure include: SO2, NO, NO2, CO, CO2, O2, H2, CH4, C2H6, C3H8, H2S, HCL, HF, NH3, CL2 from ppb ,ppm to % range.

How to ensure the accuracy and reliability of gas analyzers?
  • After the gas analyzer used some period, it needs do zeroing .After each time measurement, we should check whether the gas analyzer is back to zeroing or not , we can inlet high purity N2 backflush, And block the air inlet and outlet to ensure the accuracy of the testing data
  • Regular calibration can keep the instrument with good detection accuracy. Firstly, the instrument is calibrated with zeroing point gas and standard concentration gas, and the standard curve is obtained and stored in the gas analyzer. During measurement, the instrument compares the electrical signal generated by the gas concentration to be measured with the electrical signal of the standard concentration, and calculates the accurate gas concentration value.
  • Do not exceed the measurement concentration range. It can accurately measure the gas concentration value, but if the high-concentration gas beyond the measurement range is measured for a long time, it may cause damage to the gas sensor, affecting gas analyzer detection and life span , and even directly cause damage to the gas sensor.
  • Pay attention to temperature and humidity. Different temperature and humidity also affect the normal detection of the gas analyzer. If the temperature fluctuation in the detection environment is large, gas analyze may drift. in this case, gas analyzers can calibration is performed on-site to minimize the effect of temperature on gas detector detection results.
Do gas analyzers need to be calibrated? Why?

Generally, gas analyzers need to be calibrated to maintain high precision after being used for a period of time, but the calibration cycle is different, generally 3-6months . Our gas analyzers are equipped with an automatic zero point calibration function, which can also increase the calibration cycle. When calibrating, it is necessary to prepare the standard gas within the warranty period. Generally, the concentration of the standard gas should be selected according to the range. For some gases, gas cylinders, gas valves and pipelines should be made of anti-corrosion and anti-adsorption materials.

During calibration, the gas flow rate should be stably controlled at 1L/min- 2L/Min, or close to the actual sampling flow rate, and the flow rate should be kept stable.

What is the difference between a gas analyzer and a gas detector?

Gas analyzer and Gas detector are used to detect gas components but price big difference between the two devices , what is the difference between them?

  • different functions

The gas detector is an instrumentation tool for gas leakage concentration detection, which belongs to the safety protection instrument. A gas analyzer is an instrument used to measure the composition of a gas

  • different structure

The structure of the gas detector is relatively simple, only including the gas probe (gas sensor) and the sensor signal conversion circuit part. The gas analyzer is not only equipped with a gas sensor inside, but also has a complete set of gas circuit system including sampling system ,gas conditioning system, PLC automatic control system

  • different detection methods

The gas detector uses the probe to directly expose to the measured air or sample gas environment for detection. The gas analyzer introduces the measured gas (sample gas) into the instrument for measurement through special methods (pump sampling, in-situ sampling, etc.), and then leads it out of the instrument for emptying or recycling.

  • different detection accuracy

Gas detectors can only provide qualitative analysis results and relatively rough quantitative analysis data,A gas analyzer is a strict measuring instrument that can provide very accurate data when performing quantitative analysis.

This kind of data can be used as the basis for industrial production, gas production, safety and environmental protection improvement and improvement, and use it to guide and carry out production management, quality management and enterprise management. This kind of data can be used as an important basis for production technology, judicial appraisal, product quality supervision, scientific and technological arbitration, environmental protection emission inspection and other work.

  • Different working conditions

The gas detector does not design the adjustment and control part of the technical conditions of the sample gas, and it does not consider the environmental conditions of the sample gas, and directly detects gases. The gas analyzer internally adjusts and controls the working conditions of the sampling gas such as high temperature, high dust and moisture

  • Different methods of operation to complete the entire measurement process

When the detector is in use, just place the instrument in the measured atmosphere, and the instrument can display the measurement value. The gas analyzer must carefully introduce the sample gas into the instrument, and then strictly adjust the technical conditions of the process, such as temperature, pressure, flow, etc., only when the operator adjusts the instrument until a stable analysis of the chemical process can be obtained. Accurate measurement data.

  • Different cost

Generally speaking, the investment cost of gas detectors is low, while the cost of gas analyzers is slightly higher than gas detectors.

What is a Portable Gas Analyzer? What application does it have?

Portable gas analyzers are small, portable gas analysis instruments that are typically used in on-site detection and monitoring applications, such as environmental monitoring, industrial safety, and indoor air quality testing.

What is a continuous gas monitoring system? How is it different from conventional gas analyzers?

A continuous gas monitoring system is a system that can continuously monitor and record gas concentrations, and is usually used for long-term monitoring and automatic alarms. Compared with conventional gas analyzers, it has higher sampling frequency and data logging capability.

How do you deal with cross-interference in gas analyzers?

Cross-interference refers to the phenomenon that different gas components influence each other. To deal with cross-interference, gas analyzers typically use correction algorithms and calibration techniques to reduce or eliminate the effect of interference on measurement results.

What is spectral analysis technology? What is its application in gas analysis?

Spectroscopic analysis technology is a light-based measurement method that analyzes the characteristics of the spectrum to determine the gas composition. Common spectroscopic analysis techniques include infrared spectroscopy, ultraviolet-visible spectroscopy, and Raman spectroscopy. These techniques can be used in gas analysis to detect and measure the presence and concentration of different gases.

How does the gas analyzer collect and prepare gas samples?

The collection and preparation of gas samples can be accomplished by equipment such as sampling systems, sampling probes, and gas processing devices. Sample collection usually takes into account factors such as the selection of sampling points, sampling flow rate, and sampling time to ensure that a representative gas sample is obtained.

How does the gas analyzer perform data analysis and report generation?

Data analysis and report generation are usually performed by data processing software inside the instrument or an externally connected computer. Analysis software can process, count and graph the collected data and generate reports for further analysis and interpretation of the results.

How do gas analyzers respond to gas concentration changes and fluctuations?

Gas analyzers cope with changes and fluctuations in gas concentration by using stable sensors and calibration techniques. Calibration and automatic compensation functions help maintain the accuracy of the instrument, providing reliable measurement results even under varying conditions.

How does the gas analyzer perform multi-gas analysis?

Gas analyzers can use multiple sensors or modules to simultaneously detect and measure different gas components. Each sensor is usually specially designed to detect a specific gas, and then the measurement results of each gas are presented through the processing and display system inside the instrument.

How does the gas analyzer perform data recording and storage?

Gas analyzers usually have data recording and storage functions, and can save measurement data in internal memory or external storage devices. These data can be used for subsequent analysis, review and report generation. Some instruments also offer a data transfer feature that allows data to be transferred directly to a computer or cloud storage.

How is the power supply for the gas analyzer handled?

Gas analyzers typically have a variety of power supplies, including batteries, AC power, and DC power. Some portable instruments run on rechargeable batteries for use in mobile or field environments. Other stationary instruments may require connection to the mains or use of an external power adapter.

The response time of the gas analyzer is long, how to improve it?

The response time depends on the operating principle of the gas analyzer and the characteristics of the sensor. If the response time is long, consider using more advanced instruments or sensors to improve the sensitivity and response speed of the equipment. In addition, ensure that parameters such as flow and pressure of the sample collection and delivery system meet the requirements to speed up the gas entering the analyzer

The gas analyzer drifts, and the measurement results change with time, what should I do?

The drift of the gas analyzer may be caused by factors such as instrument aging, pollution, and stray light interference. Perform regular calibration and maintenance to clean the sensor and optical path to ensure the instrument is in top working condition. In addition, regularly check and calibrate the zero point and background value of the instrument, adjust and correct as necessary to reduce the effect of drift.

There is a difference between the measurement results displayed by the gas analyzer and the actual situation, how to solve it?

First, verify that the gas analyzer is properly calibrated and maintained. Calibration is a key step to ensure the accuracy of the instrument, you can refer to the calibration method provided by the manufacturer for operation. Also, check that the sensors are working properly and that the sample collection and handling methods are correct. If the problem persists, it may be necessary to contact the supplier for repair or replacement of the device.

What is the response time of the gas analyzer?

The response time of a gas analyzer depends on several factors, including instrument type, gas concentration, sampling system, and more. Typically, fast-response gas analyzers provide measurements within seconds, while more complex or high-precision analyzers may take minutes or longer.

What should I do if the gas analyzer shows wrong measurement results?

If the gas analyzer shows erroneous measurements, first check that it is properly calibrated and that the calibration date is not expired. If the calibration is correct and the date has not expired, maintenance and service such as cleaning the sensor, replacing consumables, or contacting the supplier for technical support may be required.

What should I do if the gas analyzer cannot detect the target gas?

If the gas analyzer cannot detect the target gas, first ensure that the concentration of the target gas is within the detection range of the instrument. If the concentration is normal and the instrument still cannot detect it, it may be necessary to check the working condition of the sensor to ensure that the sensor is not faulty or needs to be replaced. Also, check that the gas sampling system is working properly.

What should I do if the sensor of the gas analyzer responds slowly?

If your gas analyzer’s sensor responds slowly, it may be due to aging, contamination, or damage to the sensor. You can try to clean and calibrate the sensor, or contact the supplier for maintenance and sensor replacement.

How do I choose the right gas analyzer for a particular application? What factors need to be considered?
  • Confirm need gas analyzer continuous 24Hour*7 Days monitoring; measure one sampling point or inspect multiple sampling points, we can recommend you to use online or portable gas analyzer.
  • The concentration range of the measured gas and other background gases.
  • Other special function requirements, such as: output 4-20mA, RS485,RS232, or whether to use a printer or an industrial computer for data conversion, etc.
  • We will provide an on-site working condition table to understand measured gas, moisture, dust, tar, temperature, pressure, etc.
What are the types of common gas analyzers? What's the difference between them?

According to the technical principle, the commonly used is NDIR gas analyzer, UV-DOAS gas analyzer, TDLAS gas analyzer ,GC-FID gas analyzer,FTIR gas analyzer. The same gas can be measured by many different technologies. We can choose the appropriate ate gas analyzer for customers according to each customer application.

such as H2S gas

  • Due to the high analytical sensitivity of the lead acetate paper strip colorimetric method, the hydrogen sulfide content can be detected at 10×10. This type of instrument is mainly used for the monitoring of trace H2S, and the measurement range is narrow. When detecting samples with H2S content greater than 50umol/mol, the instrument needs to add a dilution system, resulting in dilution measurement errors. In addition, the lead acetate paper tape is a consumable product, which needs to be replaced frequently, and the operating cost is relatively high .
  • The ultraviolet absorption method adopts the characteristic spectral absorption principle of SO2, H2S and other gases in the ultraviolet spectral region, with high detection sensitivity and accuracy and wide measurement range. There is almost no moisture interference in the measurement of SO2, H2S and other gases by ultraviolet absorption method, which is suitable for online detection of sulfide with high humidity in the measured gas.
  • The laser optical method uses the tunable diode laser absorption spectroscopy (TDLAS) measurement technology, which is commonly used for H2S monitoring in natural gas. The working wavelength of H2S measured by laser spectroscopy is 1577nm in the near-in Frared region, and the lower detection limit can reach 5umol/mol per optical path, but it is easily interfered by gases such as CH4 and CO2, and the price of mid-infrared gas analyzers is high.
  • On-line gas chromatography Process gas chromatography can be equipped with various chromatographic detectors according to monitoring needs, so as to be suitable for on-line monitoring of different gases. It has the characteristics of many measurement components, wide measurement range, low detection limit and good repeatability. Thermal conductivity detectors (TCD) are mostly used for constant hydrogen sulfide gas detection, and flame photometric detectors (FPD) are mostly used for trace hydrogen sulfide gas detection.
  • The electrochemical sensor method is widely used in the detection and alarm of toxic and harmful gases. Electrochemical sensors are dedicated to the detection of SO2, H2S and other sulfur-containing gases. Although the measurement accuracy is not high, the price is low and easy to use. Most of them are used for portable SO2, H2S and other gas detectors, as well as for poisonous and harmful gases. Fixed detection and alarm system.
Delivery method and cycle of gas analysis equipment?

Delivery method and cycle of gas analysis equipment?

What gas components can a gas analyzer measured?

At present, the gases we can measure include: SO2, NO, NO2, CO, CO2, O2, H2, CH4, C2H6, C3H8, H2S, HCL, HF, NH3, CL2 from ppb ,ppm to % range.

How to ensure the accuracy and reliability of gas analyzers?
  • After the gas analyzer used some period, it needs do zeroing .After each time measurement, we should check whether the gas analyzer is back to zeroing or not , we can inlet high purity N2 backflush, And block the air inlet and outlet to ensure the accuracy of the testing data
  • Regular calibration can keep the instrument with good detection accuracy. Firstly, the instrument is calibrated with zeroing point gas and standard concentration gas, and the standard curve is obtained and stored in the gas analyzer. During measurement, the instrument compares the electrical signal generated by the gas concentration to be measured with the electrical signal of the standard concentration, and calculates the accurate gas concentration value.
  • Do not exceed the measurement concentration range. It can accurately measure the gas concentration value, but if the high-concentration gas beyond the measurement range is measured for a long time, it may cause damage to the gas sensor, affecting gas analyzer detection and life span , and even directly cause damage to the gas sensor.
  • Pay attention to temperature and humidity. Different temperature and humidity also affect the normal detection of the gas analyzer. If the temperature fluctuation in the detection environment is large, gas analyze may drift. in this case, gas analyzers can calibration is performed on-site to minimize the effect of temperature on gas detector detection results.
Do gas analyzers need to be calibrated? Why?

Generally, gas analyzers need to be calibrated to maintain high precision after being used for a period of time, but the calibration cycle is different, generally 3-6months . Our gas analyzers are equipped with an automatic zero point calibration function, which can also increase the calibration cycle. When calibrating, it is necessary to prepare the standard gas within the warranty period. Generally, the concentration of the standard gas should be selected according to the range. For some gases, gas cylinders, gas valves and pipelines should be made of anti-corrosion and anti-adsorption materials.

During calibration, the gas flow rate should be stably controlled at 1L/min- 2L/Min, or close to the actual sampling flow rate, and the flow rate should be kept stable.

What is the difference between a gas analyzer and a gas detector?

Gas analyzer and Gas detector are used to detect gas components but price big difference between the two devices , what is the difference between them?

  • different functions

The gas detector is an instrumentation tool for gas leakage concentration detection, which belongs to the safety protection instrument. A gas analyzer is an instrument used to measure the composition of a gas

  • different structure

The structure of the gas detector is relatively simple, only including the gas probe (gas sensor) and the sensor signal conversion circuit part. The gas analyzer is not only equipped with a gas sensor inside, but also has a complete set of gas circuit system including sampling system ,gas conditioning system, PLC automatic control system

  • different detection methods

The gas detector uses the probe to directly expose to the measured air or sample gas environment for detection. The gas analyzer introduces the measured gas (sample gas) into the instrument for measurement through special methods (pump sampling, in-situ sampling, etc.), and then leads it out of the instrument for emptying or recycling.

  • different detection accuracy

Gas detectors can only provide qualitative analysis results and relatively rough quantitative analysis data,A gas analyzer is a strict measuring instrument that can provide very accurate data when performing quantitative analysis.

This kind of data can be used as the basis for industrial production, gas production, safety and environmental protection improvement and improvement, and use it to guide and carry out production management, quality management and enterprise management. This kind of data can be used as an important basis for production technology, judicial appraisal, product quality supervision, scientific and technological arbitration, environmental protection emission inspection and other work.

  • Different working conditions

The gas detector does not design the adjustment and control part of the technical conditions of the sample gas, and it does not consider the environmental conditions of the sample gas, and directly detects gases. The gas analyzer internally adjusts and controls the working conditions of the sampling gas such as high temperature, high dust and moisture

  • Different methods of operation to complete the entire measurement process

When the detector is in use, just place the instrument in the measured atmosphere, and the instrument can display the measurement value. The gas analyzer must carefully introduce the sample gas into the instrument, and then strictly adjust the technical conditions of the process, such as temperature, pressure, flow, etc., only when the operator adjusts the instrument until a stable analysis of the chemical process can be obtained. Accurate measurement data.

  • Different cost

Generally speaking, the investment cost of gas detectors is low, while the cost of gas analyzers is slightly higher than gas detectors.

What is a Portable Gas Analyzer? What application does it have?

Portable gas analyzers are small, portable gas analysis instruments that are typically used in on-site detection and monitoring applications, such as environmental monitoring, industrial safety, and indoor air quality testing.

What is a continuous gas monitoring system? How is it different from conventional gas analyzers?

A continuous gas monitoring system is a system that can continuously monitor and record gas concentrations, and is usually used for long-term monitoring and automatic alarms. Compared with conventional gas analyzers, it has higher sampling frequency and data logging capability.

How do you deal with cross-interference in gas analyzers?

Cross-interference refers to the phenomenon that different gas components influence each other. To deal with cross-interference, gas analyzers typically use correction algorithms and calibration techniques to reduce or eliminate the effect of interference on measurement results.

What is spectral analysis technology? What is its application in gas analysis?

Spectroscopic analysis technology is a light-based measurement method that analyzes the characteristics of the spectrum to determine the gas composition. Common spectroscopic analysis techniques include infrared spectroscopy, ultraviolet-visible spectroscopy, and Raman spectroscopy. These techniques can be used in gas analysis to detect and measure the presence and concentration of different gases.

How does the gas analyzer collect and prepare gas samples?

The collection and preparation of gas samples can be accomplished by equipment such as sampling systems, sampling probes, and gas processing devices. Sample collection usually takes into account factors such as the selection of sampling points, sampling flow rate, and sampling time to ensure that a representative gas sample is obtained.

How does the gas analyzer perform data analysis and report generation?

Data analysis and report generation are usually performed by data processing software inside the instrument or an externally connected computer. Analysis software can process, count and graph the collected data and generate reports for further analysis and interpretation of the results.

How do gas analyzers respond to gas concentration changes and fluctuations?

Gas analyzers cope with changes and fluctuations in gas concentration by using stable sensors and calibration techniques. Calibration and automatic compensation functions help maintain the accuracy of the instrument, providing reliable measurement results even under varying conditions.

How does the gas analyzer perform multi-gas analysis?

Gas analyzers can use multiple sensors or modules to simultaneously detect and measure different gas components. Each sensor is usually specially designed to detect a specific gas, and then the measurement results of each gas are presented through the processing and display system inside the instrument.

How does the gas analyzer perform data recording and storage?

Gas analyzers usually have data recording and storage functions, and can save measurement data in internal memory or external storage devices. These data can be used for subsequent analysis, review and report generation. Some instruments also offer a data transfer feature that allows data to be transferred directly to a computer or cloud storage.

How is the power supply for the gas analyzer handled?

Gas analyzers typically have a variety of power supplies, including batteries, AC power, and DC power. Some portable instruments run on rechargeable batteries for use in mobile or field environments. Other stationary instruments may require connection to the mains or use of an external power adapter.

The response time of the gas analyzer is long, how to improve it?

The response time depends on the operating principle of the gas analyzer and the characteristics of the sensor. If the response time is long, consider using more advanced instruments or sensors to improve the sensitivity and response speed of the equipment. In addition, ensure that parameters such as flow and pressure of the sample collection and delivery system meet the requirements to speed up the gas entering the analyzer

The gas analyzer drifts, and the measurement results change with time, what should I do?

The drift of the gas analyzer may be caused by factors such as instrument aging, pollution, and stray light interference. Perform regular calibration and maintenance to clean the sensor and optical path to ensure the instrument is in top working condition. In addition, regularly check and calibrate the zero point and background value of the instrument, adjust and correct as necessary to reduce the effect of drift.

There is a difference between the measurement results displayed by the gas analyzer and the actual situation, how to solve it?

First, verify that the gas analyzer is properly calibrated and maintained. Calibration is a key step to ensure the accuracy of the instrument, you can refer to the calibration method provided by the manufacturer for operation. Also, check that the sensors are working properly and that the sample collection and handling methods are correct. If the problem persists, it may be necessary to contact the supplier for repair or replacement of the device.

What is the response time of the gas analyzer?

The response time of a gas analyzer depends on several factors, including instrument type, gas concentration, sampling system, and more. Typically, fast-response gas analyzers provide measurements within seconds, while more complex or high-precision analyzers may take minutes or longer.

What should I do if the gas analyzer shows wrong measurement results?

If the gas analyzer shows erroneous measurements, first check that it is properly calibrated and that the calibration date is not expired. If the calibration is correct and the date has not expired, maintenance and service such as cleaning the sensor, replacing consumables, or contacting the supplier for technical support may be required.

What should I do if the gas analyzer cannot detect the target gas?

If the gas analyzer cannot detect the target gas, first ensure that the concentration of the target gas is within the detection range of the instrument. If the concentration is normal and the instrument still cannot detect it, it may be necessary to check the working condition of the sensor to ensure that the sensor is not faulty or needs to be replaced. Also, check that the gas sampling system is working properly.

What should I do if the sensor of the gas analyzer responds slowly?

If your gas analyzer’s sensor responds slowly, it may be due to aging, contamination, or damage to the sensor. You can try to clean and calibrate the sensor, or contact the supplier for maintenance and sensor replacement.

What is the operating principle of an NDIR gas analyzer?

NDIR (Non-Dispersive Infrared) gas analyzers measure gas concentration by exploiting the property of specific gases to absorb infrared (IR) light at unique wavelengths. When IR radiation passes through a gas sample, target gas molecules absorb energy at their characteristic absorption bands. The analyzer quantifies the absorbed energy to determine the gas concentration.

Why is it called "Non-Dispersive" Infrared?

Unlike dispersive spectrometers, NDIR systems do not split light into a spectrum. Instead, they use optical filters to isolate the target gas’s absorption wavelength, simplifying the design and enhancing robustness for industrial applications.

How does the detector differentiate between gases?

Each gas has a unique IR absorption fingerprint. By pairing the detector with a narrowband optical filter, the analyzer isolates the wavelength absorbed only by the target gas (e.g., CO₂ at 4.26 μm), ensuring selectivity even in gas mixtures.

How are environmental factors like temperature compensated?

Modern NDIR analyzers integrate temperature and pressure sensors to apply real-time corrections. Advanced models also use dual-beam designs or reference channels to nullify drift caused by ambient changes or component aging.

What are the advantages of NDIR technology?

1) High specificity to target gases.
2) Long-term stability with minimal calibration drift.
3) Low maintenance due to solid-state components.
4) Wide dynamic range, suitable for ppm to percentage-level measurements.

Where is NDIR commonly applied?

NDIR analyzers are widely used in:
– Industrial emission monitoring (CO₂, CH₄, CO).
– HVAC/R systems (refrigerant leak detection).
– Environmental air quality assessment.
– Combustion efficiency optimization.

Can NDIR measure all gases?

No. NDIR is effective only for gases with IR-active molecules (diatomic gases like O₂ or N₂ cannot be measured). Common detectable gases include CO₂, CH₄, CO, SF₆, and hydrocarbons.

What is the NDIR gas analyzer?

An NDIR (Non-Dispersive Infrared) gas analyzer is a highly accurate and reliable instrument used to detect and measure the concentration of specific gases in a sample by leveraging their unique infrared (IR) absorption properties. It operates by passing infrared light through a gas sample; target gas molecules absorb specific wavelengths of IR light proportional to their concentration. A detector then quantifies the absorbed light to determine gas levels.

What is non dispersive infrared gas analysis?

Non-Dispersive Infrared (NDIR) gas analysis is a widely used optical technique for detecting and quantifying specific gases in a sample based on their unique infrared (IR) absorption properties. Unlike dispersive IR methods (e.g., FTIR), NDIR does not separate light into individual wavelengths using a prism or grating. Instead, it employs a broadband IR source, a gas sample chamber, and an optical filter to isolate the target wavelength absorbed by the gas of interest. A detector then measures the attenuated IR intensity, enabling precise concentration calculations using the Beer-Lambert law.

What is the difference between IR (Infrared) and NDIR (Non-Dispersive Infrared) technology?

1. Measurement Principle
– IR Sensors: Use broad-spectrum infrared light and may lack wavelength-specific filtering, leading to potential cross-sensitivity with non-target gases.
– NDIR Sensors: Employ a narrowband infrared source paired with optical filters to isolate specific absorption wavelengths of the target gas, minimizing interference.

2. Selectivity
– IR: Prone to interference from gases with overlapping absorption bands.
– NDIR: High selectivity due to precision optical filtering and reference/detection channel configurations.

3. Accuracy & Stability
– IR: May require frequent calibration due to environmental factors (e.g., temperature, humidity).
– NDIR: Built-in reference cells and advanced algorithms compensate for environmental drift, ensuring long-term stability (±1% accuracy typical).

4. Applications
– IR: Cost-effective for basic combustible gas detection or simple CO₂ monitoring.
– NDIR: Preferred for critical applications like industrial safety (e.g., CH₄, CO₂ leak detection), environmental monitoring (EPA compliance), and HVAC systems demanding ppm-level precision.

5. Lifespan
– IR: Shorter operational lifespan due to sensor degradation from contaminants.
– NDIR: Solid-state designs with no consumable parts often exceed 10+ years of service.

What is the difference between FID and NDIR?

1. Detection Principle
– FID (Flame Ionization Detector):
Uses a hydrogen-air flame to ionize organic compounds. The resulting ions generate a measurable current proportional to hydrocarbon concentration.
– NDIR (Nondispersive Infrared):
Measures gas concentration by detecting infrared light absorption at specific wavelengths. Gases absorb unique IR spectra, allowing selective quantification.

2. Target Gases
– FID:
Primarily detects volatile organic compounds (VOCs) and hydrocarbons (e.g., methane, propane). Insensitive to inorganic gases (e.g., CO, CO₂).
– NDIR:
Optimized for gases with strong IR absorption, including CO₂, CO, CH₄, and refrigerants. Less effective for homonuclear diatomic gases (e.g., N₂, O₂).
3. Sensitivity
– FID:
Extremely high sensitivity for hydrocarbons (ppm to ppb levels). Ideal for trace VOC analysis.
– NDIR:
Moderate sensitivity (typically ppm-level). Performance depends on gas-specific absorption strength.

4. Interference & Selectivity
– FID:
Broadly responds to most hydrocarbons but cannot distinguish between them. Requires chromatographic separation for speciated analysis.
– NDIR:
Highly selective due to wavelength-specific filters. Minimal cross-interference when properly configured.
5. Maintenance & Operational Requirements
– FID:
Requires hydrogen fuel gas, regular flame maintenance, and frequent calibration.
– NDIR:
No consumables (e.g., fuel). Maintenance focuses on optical cleanliness and periodic calibration.
6. Typical Applications
– FID:
Environmental monitoring (VOC emissions), industrial process control (refineries), and gas chromatography.
– NDIR:
Combustion analysis (CO₂, CO), indoor air quality monitoring, automotive emissions testing, and HVAC systems.

What is the fundamental difference between dispersive and non-dispersive systems?

Dispersive Systems – Definition: Exhibit frequency-dependent phase velocity, causing waves of different frequencies to travel at different speeds.
– Physical Manifestation: Produces chromatic dispersion (in optics) or frequency dispersion (in acoustics/mechanical waves).
– Examples:
Prism-based spectrometers (optical dispersion)
Multi-mode optical fibers
Surface acoustic wave (SAW) devices with frequency-dependent delay
– Key Feature: Wavelength separation or pulse broadening over propagation distance.

Non-Dispersive Systems
– Definition: Maintain frequency-independent phase velocity, preserving wave shape during propagation.
– Physical Behavior: All frequency components propagate at identical speeds (no velocity spread).
– Examples:
Ideal transmission lines (TEM mode)
Non-dispersive infrared (NDIR) gas sensors using fixed-wavelength detection
Vacuum electromagnetic wave propagation
– Key Feature: Minimal signal distortion and temporal spreading.

What are the key limitations or disadvantages of NDIR (Non-Dispersive Infrared) technology?

While NDIR is widely used for gas detection (e.g., CO₂, hydrocarbons), it has several inherent limitations:

1. Cross-Sensitivity Issues: NDIR sensors may suffer from interference when multiple gases have overlapping infrared absorption bands (e.g., methane and water vapor), requiring advanced filtering or compensation algorithms.

2. High Cost: Precision optical components (e.g., infrared sources, detectors, and filters) increase manufacturing costs compared to electrochemical or catalytic bead sensors.

3. Limited Sensitivity for Low Concentrations: NDIR struggles to detect trace gas levels (e.g., sub-ppm for VOCs) due to weak absorption signals, making it less suitable for applications requiring ultra-low detection limits.

4. Temperature and Pressure Dependence: Sensor accuracy can drift with ambient temperature or pressure fluctuations, necessitating built-in compensation mechanisms.

5. Maintenance Requirements: Optical windows are prone to contamination (e.g., dust, condensation), leading to calibration drift and requiring periodic cleaning or replacement.

6. Power Consumption: Continuous operation of infrared sources (e.g., microheaters) results in higher power demand, limiting battery-powered deployment.

7. Slow Response Time: NDIR typically has slower response times (seconds to minutes) compared to technologies like photoionization detectors (PID), hindering real-time monitoring in dynamic environments.

8. Limited Multi-Gas Capability: Simultaneous detection of multiple gases often requires separate optical channels, increasing system complexity and cost.

What is the difference between dispersive and non-dispersive IR spectrometers?

1. Optical Design:
– Dispersive IR Spectrometers: Use a monochromator (e.g., prism or diffraction grating) to physically separate infrared wavelengths. Light is dispersed spatially, and a detector scans across the spectrum.
– Non-Dispersive IR (NDIR) Spectrometers: Lack a monochromator. Instead, they employ optical filters or gas-filled cells to isolate specific wavelengths, often paired with a broadband detector.

2. Resolution & Spectral Range:
– Dispersive: High spectral resolution (0.1–4 cm⁻¹), ideal for detailed molecular fingerprinting across a broad IR range (e.g., 400–4000 cm⁻¹).
– NDIR: Limited to pre-selected wavelengths (e.g., CO₂ at 4.26 µm), optimized for targeted gas detection with minimal spectral interference.

3. Mechanical Complexity:
– Dispersive: Requires moving parts (e.g., rotating grating), increasing maintenance needs and sensitivity to vibration.
– NDIR: Solid-state design with no moving parts, enhancing ruggedness and reliability for field/industrial use.

4. Applications:
– Dispersive: Research-grade qualitative analysis (e.g., identifying unknown compounds, studying molecular structure).
– NDIR: Quantitative monitoring of specific gases (e.g., CO₂ in emissions, methane in leak detection) with high sensitivity and real-time response.

5. Cost & Speed:
– Dispersive: Higher cost, slower scanning due to sequential wavelength measurement.
– NDIR: Lower cost, faster response (milliseconds), suited for continuous monitoring.

What is the principle of a Non-Dispersive Infrared (NDIR) sensor?

A Non-Dispersive Infrared (NDIR) sensor operates based on the principle of infrared light absorption by gas molecules. Specific gases absorb infrared (IR) radiation at unique wavelengths due to their molecular structure. The sensor uses an infrared light source, an optical filter (to isolate the target gas’s absorption wavelength), and a photodetector to measure the intensity of transmitted light. Gas concentration is calculated by comparing absorbed vs. transmitted IR energy, following the Beer-Lambert Law.

How does an ultrasonic flow meter work?

An ultrasonic flow meter measures fluid flow velocity using high-frequency sound waves. It operates based on two primary principles: transit-time differential and Doppler effect, depending on the fluid type and application.

1. Transit-Time Method (Time-of-Flight):
– Two ultrasonic transducers (sensors) are mounted on the pipe, either in a clamp-on (non-invasive) or wetted (invasive) configuration.
– The sensors alternately transmit and receive ultrasonic pulses upstream and downstream through the fluid.
– The difference in transit time (Δt) between the two directions is measured. Faster-moving fluids shorten the upstream pulse time and lengthen the downstream pulse time.

2. Doppler Effect Method:
– Suitable for fluids with suspended particles or bubbles (e.g., wastewater, slurries).
– A single transducer emits ultrasonic waves, which reflect off moving particles in the fluid.
– The frequency shift (Doppler shift) between the transmitted and reflected waves is proportional to the fluid velocity.

What is the basic operating principle of an ultrasonic gas flow meter?

Ultrasonic gas flow meters measure flow velocity by transmitting high-frequency sound waves through the gas stream. They calculate flow rate by analyzing the time difference (transit time difference) between ultrasonic signals traveling with the flow (downstream) and against the flow (upstream). This time difference is directly proportional to the gas velocity.

What components are critical to an ultrasonic gas flow meters' operation?

Key components include:
1. Ultrasonic Transducers: Paired sensors that alternately transmit and receive ultrasonic pulses.
2. Signal Processors: Measure transit times and convert time differences into velocity data.
3. Temperature/Pressure Sensors: Compensate for gas density changes to ensure volumetric or mass flow accuracy.
4. Flow Calculator: Integrates velocity, pipe cross-section area, and gas properties to compute flow rate.

What are the disadvantages of an ultrasonic gas flow meter?

While ultrasonic gas flow meters offer advantages like non-intrusive measurement and high accuracy, they also have limitations. Key disadvantages include:

1. Sensitivity to Flow Profile Disturbances: Requires sufficient straight pipe runs upstream/downstream to stabilize flow profiles. Irregularities (e.g., bends, valves) can cause measurement errors.
2. High Cost: Advanced models with high accuracy and diagnostics are expensive compared to conventional meters (e.g., diaphragm, turbine).
3. Limited Performance in Dirty Gases: Particulates, moisture, or heavy contaminants can attenuate ultrasonic signals, reducing reliability.
4. Temperature and Pressure Dependencies: Extreme temperature/pressure variations may affect speed-of-sound calculations, requiring compensation.
5. Lower Accuracy at Low Flow Rates: Signal-to-noise ratio diminishes in low-velocity flows, increasing uncertainty.
6. Complex Installation and Calibration: Proper alignment of transducers is critical; improper installation leads to drift or failure.
7. Susceptibility to Acoustic Noise: External vibrations or ultrasonic interference (e.g., from machinery) may disrupt measurements.

What parameters or properties can ultrasonic sensors and systems measure?

1.Distance/Position: Ultrasonic sensors calculate distance by emitting high-frequency sound waves and measuring the time delay (time-of-flight) of the reflected echo. Applications include object detection, liquid level monitoring, and parking assistance systems.

2. Flow Rate: Ultrasonic flow meters use the *Doppler effect* or *transit-time difference* to measure the velocity of liquids or gases in pipelines, enabling non-invasive flow rate calculations.

3. Thickness: Ultrasonic thickness gauges measure material thickness (e.g., metal, plastic, glass) by analyzing the time taken for sound waves to travel through a material and reflect from its back surface.

4. Structural Integrity: Ultrasonic testing (UT) detects flaws (cracks, voids, corrosion) in materials by identifying changes in wave propagation, attenuation, or reflection patterns.

5. Material Properties: Ultrasonic waves can characterize material properties such as density, elasticity, and homogeneity by analyzing wave speed, absorption, and scattering.

6. Presence/Absence: Used in industrial automation, ultrasonic sensors detect the presence or absence of objects without physical contact.

What are the advantages of ultrasonic gas flow meters?

Non-intrusive Design: No moving parts or pressure drop.
– Bidirectional Flow Measurement: Detects forward and reverse flow.
– Wide Turndown Ratio: Accurate across a broad flow range (e.g., 1:100).
– Low Maintenance: Immune to contamination or wear.
– Large Pipe Compatibility: Effective for diameters from 0.5″ to over 120″.

In which applications are ultrasonic gas flow meters preferred?

1) Natural gas distribution and custody transfer.
2) Emissions monitoring (e.g., flare gas measurement).
3) Compressed air systems and biogas plants.
4) High-pressure or corrosive gas environments.

What is the primary function of an ultrasonic Doppler flow detector?

An ultrasonic Doppler flow detector measures the velocity and volumetric flow rate of liquids or gases in a closed conduit (e.g., pipes, ducts) using the Doppler effect. It is specifically designed for fluids containing suspended particles, bubbles, or inhomogeneities that reflect ultrasonic waves.

How does an ultrasonic Doppler flow detector work?

The device emits high-frequency ultrasound waves (typically 0.5–10 MHz) into the fluid via a transducer. Moving particles or bubbles in the flow scatter the waves, causing a Doppler shift (frequency change) proportional to the fluid velocity. The detector analyzes this shift to calculate flow velocity and derives volumetric flow rate using the pipe’s cross-sectional area.

What is the typical accuracy range of an ultrasonic gas flow meter?

Ultrasonic gas flow meters typically achieve ±0.5% to ±1% of reading under ideal conditions, depending on the model, measurement principle (transit-time or Doppler), and installation quality. Advanced meters with high-precision calibration and stable flow profiles can reach accuracies as high as ±0.3%.

What factors influence the accuracy of ultrasonic gas flow meters?

1. Flow profile stability: Turbulence or uneven flow distribution reduces accuracy.
2. Gas composition: Changes in density, viscosity, or impurities (e.g., particulates) affect signal clarity.
3. Temperature and pressure variations: Most meters require real-time compensation using integrated sensors.
4. Installation quality: Proper alignment, sufficient straight pipe runs (typically 10D upstream/5D downstream), and avoiding vibrations are critical.
5. Sensor fouling: Contamination on transducer surfaces degrades performance over time.

How do clamp-on vs. inline ultrasonic meters compare in accuracy?

Inline meters (wetted transducers) generally offer higher accuracy (±0.5–1%) due to direct signal transmission through the gas. Clamp-on meters (non-invasive) may have slightly reduced accuracy (±1–2%) but are ideal for retrofitting or hazardous environments.

How can accuracy be maintained long-term?

Regular field verification (e.g., with portable reference meters) and recalibration every 1–3 years are recommended. Self-diagnostic features in modern meters (e.g., signal quality indicators) help detect drift early.

What is the typical lifespan of an ultrasonic gas flow meter?

The operational lifespan of an ultrasonic gas flow meter generally ranges between 5 to 15 years, depending on critical factors such as product quality, environmental conditions, and maintenance practices. High-quality meters with corrosion-resistant sensors and robust electronic components, when installed in controlled environments (e.g., moderate temperature, minimal vibration, and non-corrosive media), can achieve the upper end of this range. In contrast, units exposed to harsh conditions (e.g., high pressure, corrosive gases, or excessive dust) may experience reduced longevity.

Proactive maintenance—including regular sensor calibration, cable integrity checks, and debris removal—significantly extends service life. Advanced models with redundant measurement channels or enhanced filtration systems (e.g., integrated particulate filters) further improve durability. For example, dual-channel designs allow continuous operation even if one sensor fails, while filtration mitigates damage from contaminants.
Key components like transducers typically last 8–10 years, whereas electronic modules (e.g., transmitters) may function reliably for 12–15 years under optimal conditions. Always adhere to manufacturer guidelines for installation and operational limits (e.g., pressure, temperature) to maximize performance and lifespan.

Why does my ultrasonic gas flow meter trigger false alarms, and how can I prevent them?

Unnecessary alarms in ultrasonic gas flow meters are typically caused by improper installation, environmental interference, or configuration issues. Below are common causes and professional solutions:

1. Incorrect Installation
– Cause: Insufficient straight pipe lengths upstream/downstream, or obstructions (e.g., valves, bends) disrupting flow profile.
– Solution: Follow manufacturer guidelines for minimum straight pipe requirements (typically 10D upstream and 5D downstream, where D = pipe diameter). Ensure sensors are aligned precisely and mounted securely.

2. Environmental Interference
– Cause: Temperature fluctuations, vibration, or electromagnetic noise affecting signal integrity.
– Solution:
– Stabilize ambient temperature and isolate the meter from excessive vibration.
– Use shielded cables and proper grounding to mitigate electromagnetic interference (EMI).

3. Contaminated Sensors or Pipe Walls
– Cause: Buildup of debris, moisture, or condensate on transducers or pipe surfaces.
– Solution: Install filters or moisture separators upstream. Schedule regular maintenance for sensor cleaning and inspect pipe integrity.

4. Incorrect Parameter Settings
– Cause: Overly sensitive alarm thresholds or mismatched gas properties (e.g., density, composition).
– Solution:
– Recalibrate the meter for the specific gas composition and operating conditions.
– Adjust alarm thresholds (e.g., flow rate limits, signal quality thresholds) based on historical data.

5. Acoustic Signal Degradation
– Cause: Attenuation due to high gas velocity, excessive turbulence, or incompatible gas mixtures.
– Solution: Verify the meter is rated for the gas type and velocity range. Optimize signal processing settings (e.g., gain, signal-to-noise ratio).

6. Power Supply Issues
– Cause: Voltage fluctuations or poor grounding.
– Solution: Use a stabilized power supply and ensure proper grounding per IEC/ISA standards.

Pro Tip: Perform routine diagnostics using the meter’s built-in software to monitor signal quality (e.g., SNR values) and validate transducer performance. For persistent issues, consult the manufacturer’s technical support for firmware updates or advanced troubleshooting.

What steps can I take to maintain the accuracy and reliability of an ultrasonic gas flow meter under unstable power supply conditions?

To ensure stable operation of ultrasonic gas flow meters in environments with significant power supply fluctuations, implement the following industry-recommended practices:

1. Use a Voltage Regulator/Stabilizer
Deploy a high-quality voltage regulator or uninterruptible power supply (UPS) to mitigate input voltage fluctuations. This ensures the meter receives a consistent voltage (e.g., 24V DC or 120/230V AC) within its specified tolerance range (±10% typically).

2. Install Power Conditioning Filters
Integrate EMI/RFI filters or surge protectors to suppress electrical noise, harmonics, and transient voltage spikes that may interfere with the meter’s signal processing or damage sensitive components.

3. Select Models with Wide Input Voltage Ranges
Opt for flow meters designed for industrial-grade power compatibility (e.g., 9–36V DC or 85–265V AC). These models often include built-in voltage regulation and transient protection.

4. Ensure Proper Grounding and Shielding
Follow IEC 61000 standards for grounding to eliminate ground loops and shield cables to reduce electromagnetic interference (EMI) from affecting power integrity.

5. Verify Power Supply Redundancy
For critical applications, use redundant power supplies (dual DC inputs or backup batteries) to prevent downtime during power interruptions.

6. Conduct Regular Power Quality Audits
Monitor voltage, current, and frequency stability using power quality analyzers to identify and address anomalies before they impact meter performance.

7. Leverage Low-Power Operating Modes
Activate sleep modes or low-power algorithms (if supported) to reduce energy consumption during voltage dips without compromising measurement continuity.

What measures can be taken to protect flowmeters from strong magnetic field interference?

To minimize or eliminate magnetic field interference on flowmeters, implement the following industry-recommended strategies:

1. Select Magnetically Robust Designs
– Opt for flowmeters with EMC (Electromagnetic Compatibility) certification or those specifically designed for high-magnetic environments (e.g., pulsed DC electromagnetic flowmeters with noise suppression).
– Avoid using devices with unshielded analog signal outputs in areas with strong magnetic fields.

2. Maintain Safe Distance from Interference Sources
– Install the flowmeter ≥3 meters (10 feet) away from high-power equipment (e.g., transformers, motors, VFDs) to reduce magnetic flux density.
– Follow the inverse-square law: doubling the distance from a magnetic source reduces interference by ~75%.

3. **Implement Magnetic Shielding**
– Enclose the flowmeter and/or cabling in Mu-metal (high-permeability alloy) or ferromagnetic enclosure to redirect magnetic field lines.
– Use twisted-pair or coaxial cables with braided shielding grounded at a single point to prevent ground loops.

4. Optimize Grounding Practices
– Establish a dedicated grounding system (≤1Ω resistance) separate from power grounds to avoid induced currents.
– Use galvanic isolation for signal lines to block conductive interference paths.

5. Apply Signal Filtering
– Integrate low-pass filters (e.g., RC filters) or digital signal processing (DSP) algorithms to attenuate high-frequency magnetic noise.
– For analog outputs, use 4-20mA HART® or Foundation Fieldbus™ protocols with inherent noise immunity.

6. Validate Installation via Testing
– Perform EMI/RFI scans before installation to identify ambient magnetic field levels.
– Post-installation, conduct zero-point calibration under no-flow conditions to detect residual interference.

7. Consult Manufacturer Guidelines
– Adhere to the flowmeter manufacturer’s installation manual for orientation, shielding requirements, and compatibility with IEC 61326-1 (EMC standards for industrial equipment).

Get An Quote

We’ll send you the catalog as soon as you submit your email